- Climatic Chambers
StabilClima 600 Climatic Chamber
Ideal for photostability testing in pharmaceuticals, cosmetics, and equivalents, compliant with ICH Q1B (Option 2) guidelines. Features integrated UV and Visible light sensors for precise automatic management. Its wide climatic range also allows for climatic stability testing (ICH Q1A). Aralab’s choice for precision, compliance, and versatility.

Description
An environmental chamber for photostability testing fully compliant with ICH guidelines, performing UV and Visible light tests in under 50 hours, with accumulated light emission tracking and automatic stop function. With a wide climatic range and adaptable interior, this chamber offers great versatility, also enabling the execution of climatic stability tests (ICH Q1A), optimizing your investment. The controller and software are compliant with FDA 21 CFR part 11 for secure data logging. We provide all IQ, OQ, and PQ (Installation, Operational, Performance Qualification) documents, essential for your validation and quality requirements.
Specifications
- INNER DIMENSIONS (H X W X D) (MM)1340 X 600 X 660
- EXTERNAL DIMENSIONS (H X W X D) (MM)1980 X 730 X 920
- INTERNAL VOLUME (LITERS)543
- TEMPERATURE RANGE-5ºC TO 45ºC (Lights Off) +5ºC TO 45ºC (Lights On)
- HUMIDITY RANGE (rH%)35 TO 95% rH (Lights Off) 40 TO 80% rH (Lights On)
- VISIBLE LIGHT26,000 LUX (+/- 6% APPROX.) TEST ICH Q1B VISIBLE IN APPROXIMATELY 48H
- UV LIGHT30 W/M² (+/- 10% APPROX.) TEST ICH Q1B UV IN APPROXIMATELY 7H
- SHELVES / STORAGE (M²)ONE SHELF WITH VISIBLE LIGHTS (0.14M²) ONE SHELF WITH UV LIGHTS (0.14M²)
- Available sizes
- 543L

Product highlights
Product features
Explore our chamber features.
custom solutions